Device recall · November 16, 2022
Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711 recall
Why: Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Insulin Pump/Model: MiniMed 620G/ MMT-1750 MiniMed 640G/ MMT-1711, MMT-1712, MMT-1751, MMT-1752
- Lots / codes
Model/UDI-GTIN (All Serial Numbers): MMT-1750/00763000375461, 00643169559745; MMT-1711/643169554931, 763000367039, 00643169890039, 00643169554917, 00643169629813, 00643169554924, 00643169554931, 00643169742062, 00763000013066, 00643169554948, 00643169554955, 00643169621954; MMT-1712/643169662612, 643169577701, 00763000205409, 00643169890046, 00643169577664, 00643169629820, 00643169577671, 00643169577688, 00763000155346, 00643169577695, 00643169577701, 00643169621961; MMT-1751/643169672239, 00643169574410, 00643169574427, 00643169574434, 00763000253288, 00643169574441, 00643169574458, 00643169643512, 00643169521155, 00643169521704, 00763000015596, 00643169520882, 00643169521421, 00643169521971; MMT-1752/763000192181, 00643169596368, 00643169596382, 00643169596405, 00763000318291, 00763000318307, 00643169596443, 00643169522305, 00643169521292, 00643169522381, 00643169521841, 00643169522268, 00643169521025, 00643169522343, 00643169521575, 00643169522114
- Quantity
- 316326
- Company
- Medtronic MiniMed, Northridge, CA
- Distribution
- US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela
- Initiated
- September 20, 2022 · Voluntary: Firm initiated
- Classified
- November 7, 2022
- Reported
- November 16, 2022
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.