Device recall · November 23, 2022
18239: ASM Omnipod Dash PDM recall
Why: The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.
- Lots / codes
Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots
- Quantity
- 248,288 US; 167,802 OUS
- Company
- Insulet Corporation, Acton, MA
- Distribution
- Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.
- Initiated
- October 17, 2022 · Voluntary: Firm initiated
- Classified
- November 15, 2022
- Reported
- November 23, 2022
- Ended
- September 17, 2024
- How notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall