Class IOngoingZ-0151-2024
Device recall · November 15, 2023
Novum IQ Syringe infusion system recall
Why: Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Novum IQ Syringe infusion system, Product Code 40800BAXUS
- Lots / codes
All Serial Numbers
- Quantity
- 2023 units
- Company
- Baxter Healthcare Corporation, Deerfield, IL
- Distribution
- US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
- Initiated
- October 13, 2023 · Voluntary: Firm initiated
- Classified
- November 9, 2023
- Reported
- November 15, 2023
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.