Device recall · October 26, 2022
Medicina ENFit 1ml Reusable Low Dose Enteral Syringe Code: LHE01LD recall
Why: Plunger component has been manufactured with an unapproved material (Polyethylene not Polypropylene) and does not meet raw material specification
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Medicina ENFit 1ml Reusable Low Dose Enteral Syringe Code: LHE01LD
- Lots / codes
UDI-DI: (01)05060278508146 Batch number: HM17121, HM18051, HM18080, HM18120, HM18141, HM19007, HM19016, HM19059, HM19083, HM19130, HM19144, HM20005, HM20054, HM21044, HM22018
- Quantity
- 9 boxes (900 syringes)
- Company
- Medicina Uk Ltd, Bolton, N/A
- Distribution
- Nationwide
- Initiated
- July 4, 2022 · Voluntary: Firm initiated
- Classified
- October 20, 2022
- Reported
- October 26, 2022
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.