Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IIOngoingZ-0080-2024

Device recall · October 18, 2023

Medfusion PLUNGER FLOAT PLATE recall

Why: A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

Medfusion PLUNGER FLOAT PLATE, Component Number G6000069, used on infusion pump

Lots / codes

Lot Numbers: P0048777, P0067276, P0094724, P0102958, P0131089, P0266366, P0412271, P0725712, P0729657, P0744687, P0745394, P112621, P122904, P129485

Quantity
674 units
Company
Smiths Medical ASD Inc., Minneapolis, MN
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
August 16, 2023 · Voluntary: Firm initiated
Classified
October 6, 2023
Reported
October 18, 2023
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Recalled with it (same recall event)

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.

Every price you share shows someone else what's fair - and which pharmacies charge the least. It takes 30 seconds and it's anonymous.