Device recall · October 9, 2024
MiniMed 630G Insulin Pump recall
Why: Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851
- Lots / codes
REF/UDI-DI(GTIN): MMT-1714/00643169782396, 00763000253059; MMT-1715/00643169873841, 00763000190446, 00643169752726, 643169873834, 00763000317591, 00763000316655, 763000367053, 763000316655, 76300031665503, 763000190439, 763000316631, 763000166519, 763000072520, 76300031665501, 763000256807, 763000090197, 643169656840, 763000316686, 00643169656840, 00763000090197, 00763000190439, 00763000316631; MMT-1755/643169752726, 643169873841, 763000190446, 763000101527, 763000317591, 00763000317591, 763000367091, 763000179632, 763000317621, 643169878648, 763000157043, 76300031759103, 00643169752726, 76300031759101, 76300031762103, 76300031762101, 00763000101527, 00763000190446; MMT-1801/00763000531157; MMT-1805/00763000561680; MMT-1850/00763000552541; MMT-1851/00763000854737, 00763000541323. All serial numbers.
- Quantity
- 24,595
- Company
- Medtronic MiniMed, Inc., Northridge, CA
- Distribution
- US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide
- Initiated
- July 31, 2024 · Voluntary: Firm initiated
- Classified
- October 3, 2024
- Reported
- October 9, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.