Drug recall · October 16, 2019
Sodium Chloride Injection USP 0.9% recall
Why: Labeling: Not Elsewhere Classified: RFID formatting error which reads product in Ultraject syringe as expired and will not allow injection of product by the Optivantage injector.
Class III: unlikely to cause harm, but breaks the rules (for example a labeling error).
The recall
- Product
Sodium Chloride Injection USP 0.9%, packaged in 1 - 125 mL Ultraject Prefilled Syringe For Power Injection per carton, RFID Technology, Rx Only, Manufactured by: Liebel-Flarshelm Company, LLC., Raleigh, NC 27616; NDC 0019-1188-27.
- Lots / codes
Lots: L159A, Exp. 06/2021; L169A, Exp. 07/2021
- Quantity
- 3140 syringes
- Company
- Guerbet LLC, Princeton, NJ
- Distribution
- Nationwide in the USA
- Initiated
- September 13, 2019 · Voluntary: Firm initiated
- Classified
- October 9, 2019
- Reported
- October 16, 2019
- Ended
- October 25, 2021
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)